<?xml version="1.0" encoding="UTF-8"?>
<!DOCTYPE article PUBLIC "-//TaxonX//DTD Taxonomic Treatment Publishing DTD v0 20100105//EN" "https://pharmacia.pensoft.net/nlm/tax-treatment-NS0.dtd">
<article xmlns:tp="http://www.plazi.org/taxpub" xmlns:xsi="http://www.w3.org/2001/XMLSchema-instance" xmlns:xlink="http://www.w3.org/1999/xlink" xmlns:mml="http://www.w3.org/1998/Math/MathML" article-type="research-article" xml:lang="en">
  <front>
    <journal-meta>
      <journal-id journal-id-type="publisher-id">89</journal-id>
      <journal-id journal-id-type="index">urn:lsid:arphahub.com:pub:CCBBD2E1-BDAF-5702-AE4B-0E4910A4C7D6</journal-id>
      <journal-title-group>
        <journal-title xml:lang="en">Pharmacia</journal-title>
        <abbrev-journal-title xml:lang="en">PHAR</abbrev-journal-title>
      </journal-title-group>
      <issn pub-type="ppub">0428-0296</issn>
      <issn pub-type="epub">2603-557X</issn>
      <publisher>
        <publisher-name>Bulgarian Pharmaceutical Scientific Society</publisher-name>
      </publisher>
    </journal-meta>
    <article-meta>
      <article-id pub-id-type="doi">10.3897/pharmacia.73.e200063</article-id>
      <article-id pub-id-type="publisher-id">200063</article-id>
      <article-categories>
        <subj-group subj-group-type="heading">
          <subject>Research Article</subject>
        </subj-group>
        <subj-group subj-group-type="scientific_subject">
          <subject>Clinical pharmacology</subject>
        </subj-group>
      </article-categories>
      <title-group>
        <article-title>A comparative study of visual outcomes and patient satisfaction with Evolux™, monofocal, and extended depth of focus intraocular lenses</article-title>
      </title-group>
      <contrib-group content-type="authors">
        <contrib contrib-type="author" corresp="no">
          <name name-style="western">
            <surname>Zdravkov</surname>
            <given-names>Yani</given-names>
          </name>
          <xref ref-type="aff" rid="A1">1</xref>
          <xref ref-type="aff" rid="A2">2</xref>
        </contrib>
        <contrib contrib-type="author" corresp="yes">
          <name name-style="western">
            <surname>Dzhambazov</surname>
            <given-names>Radomir</given-names>
          </name>
          <email xlink:type="simple">radodzhambazov@gmail.com</email>
          <xref ref-type="aff" rid="A1">1</xref>
          <xref ref-type="aff" rid="A2">2</xref>
        </contrib>
        <contrib contrib-type="author" corresp="no">
          <name name-style="western">
            <surname>Veleva</surname>
            <given-names>Nevyana</given-names>
          </name>
          <uri content-type="orcid">https://orcid.org/0000-0002-8055-3451</uri>
          <xref ref-type="aff" rid="A1">1</xref>
          <xref ref-type="aff" rid="A2">2</xref>
        </contrib>
        <contrib contrib-type="author" corresp="no">
          <name name-style="western">
            <surname>Haykin</surname>
            <given-names>Vasil</given-names>
          </name>
          <uri content-type="orcid">https://orcid.org/0000-0001-8539-8860</uri>
          <xref ref-type="aff" rid="A1">1</xref>
          <xref ref-type="aff" rid="A2">2</xref>
        </contrib>
        <contrib contrib-type="author" corresp="no">
          <name name-style="western">
            <surname>Oscar</surname>
            <given-names>Alexander</given-names>
          </name>
          <uri content-type="orcid">https://orcid.org/0000-0003-3492-9531</uri>
          <xref ref-type="aff" rid="A1">1</xref>
          <xref ref-type="aff" rid="A2">2</xref>
        </contrib>
      </contrib-group>
      <aff id="A1">
        <label>1</label>
        <addr-line content-type="verbatim">Eye Clinic, Alexandrovska University Hospital, Sofia, Bulgaria</addr-line>
        <institution>Eye Clinic, Alexandrovska University Hospital</institution>
        <addr-line content-type="city">Sofia</addr-line>
        <country>Bulgaria</country>
        <uri content-type="ror">https://ror.org/04b8y3f13</uri>
      </aff>
      <aff id="A2">
        <label>2</label>
        <addr-line content-type="verbatim">Department of Ophthalmology, Medical University of Sofia, Sofia, Bulgaria</addr-line>
        <institution>Department of Ophthalmology, Medical University of Sofia</institution>
        <addr-line content-type="city">Sofia</addr-line>
        <country>Bulgaria</country>
        <uri content-type="ror">https://ror.org/01n9zy652</uri>
      </aff>
      <author-notes>
        <fn fn-type="corresp">
          <p>Corresponding author: Radomir Dzhambazov (<email xlink:type="simple">radodzhambazov@gmail.com</email>)</p>
        </fn>
      </author-notes>
      <pub-date pub-type="collection">
        <year>2026</year>
      </pub-date>
      <pub-date pub-type="epub">
        <day>25</day>
        <month>09</month>
        <year>2026</year>
      </pub-date>
      <volume>73</volume>
      <elocation-id>e200063</elocation-id>
      <uri content-type="arpha" xlink:href="http://openbiodiv.net/1278F877-5CE6-5A3B-B164-E199DCCF298B">1278F877-5CE6-5A3B-B164-E199DCCF298B</uri>
      <history>
        <date date-type="received">
          <day>17</day>
          <month>05</month>
          <year>2026</year>
        </date>
        <date date-type="accepted">
          <day>01</day>
          <month>09</month>
          <year>2026</year>
        </date>
      </history>
      <permissions>
        <copyright-statement>Yani Zdravkov, Radomir Dzhambazov, Nevyana Veleva, Vasil Haykin, Alexander Oscar</copyright-statement>
        <license license-type="creative-commons-attribution" xlink:href="http://creativecommons.org/licenses/by/4.0/" xlink:type="simple">
          <license-p>This is an open access article distributed under the terms of the Creative Commons Attribution License (CC BY 4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.</license-p>
        </license>
      </permissions>
      <abstract>
        <label>Abstract</label>
        <p>This retrospective study aimed to evaluate the postoperative visual acuity and overall patient satisfaction of three groups of patients who underwent cataract surgery. A total of 56 patients and 67 eyes were evaluated and divided based on the type of intraocular lens (<abbrev xlink:title="intraocular lens">IOL</abbrev>) implanted. The evaluation consisted of examining visual acuity at far and intermediate distances preoperatively and postoperatively. On the 30<sup>th</sup> postoperative day, contrast sensitivity was evaluated using the Pelli-Robson chart, and all patients completed a questionnaire addressing patient satisfaction at far and intermediate distances, dependence on spectacles, occurrence of halos and glare, and driving difficulties. The present study highlights a clinically relevant intermediate vision advantage of the Evolux™ enhanced monofocal <abbrev xlink:title="intraocular lens">IOL</abbrev> over a standard monofocal <abbrev xlink:title="intraocular lens">IOL</abbrev> while maintaining comparable distance visual acuity and contrast sensitivity. Although extended depth of focus (<abbrev xlink:title="extended depth of focus">EDOF</abbrev>) <abbrev xlink:title="intraocular lenses">IOLs</abbrev> provided superior intermediate visual acuity, they were associated with lower contrast sensitivity and more frequent photic phenomena in this cohort. The study highlights that enhanced monofocal <abbrev xlink:title="intraocular lenses">IOLs</abbrev> may represent a useful compromise between the limited intermediate vision of conventional monofocal lenses and the optical trade-offs associated with <abbrev xlink:title="extended depth of focus">EDOF</abbrev> lenses. Larger prospective studies with standardized patient-reported outcome measures and longer follow-up are warranted.</p>
      </abstract>
      <kwd-group>
        <label>Keywords</label>
        <kwd>cataract</kwd>
        <kwd>Evolux™</kwd>
        <kwd>extended depth of focus</kwd>
        <kwd>intraocular lens</kwd>
        <kwd>monofocal</kwd>
      </kwd-group>
    </article-meta>
  </front>
  <body>
    <sec sec-type="Introduction" id="sec1">
      <title>Introduction</title>
      <p>Cataract surgery with the implantation of an <abbrev xlink:title="intraocular lens">IOL</abbrev> is the most commonly performed surgery in ophthalmology and the only treatment for lens opacification (<xref ref-type="bibr" rid="B8">Moshirfar et al. 2023</xref>; <xref ref-type="bibr" rid="B7">Li et al. 2024</xref>).</p>
      <p>In the era of constantly increasing dependence on electronic devices, such as smartphones and tablets, the demand for clear intermediate and near vision has increased. Therefore, manufacturers are working toward offering a wide variety of <abbrev xlink:title="intraocular lenses">IOLs</abbrev> to suit patients’ needs at affordable prices (<xref ref-type="bibr" rid="B2">Dell et al. 2024</xref>). With increased access to information, patients with good visual acuity are seeking to undergo cataract surgery. These patients often have a demanding lifestyle and high expectations for their postoperative visual acuity at various distances. It is important to discuss the different <abbrev xlink:title="intraocular lens">IOL</abbrev> options and the expected outcomes in terms of the use of spectacles. Moreover, phacoemulsification is becoming increasingly popular among younger individuals who have not developed a cataract but are seeking to achieve emmetropia with minimal use of spectacles. Enhanced monofocal <abbrev xlink:title="intraocular lenses">IOLs</abbrev> such as Evolux™ may be a suitable choice for these patients in terms of visual outcome and cost (<xref ref-type="bibr" rid="B2">Dell et al. 2024</xref>).</p>
      <p>Standard monofocal lenses provide a fixed focus at one distance. They can be calculated to achieve a focus point for either far or near vision (<xref ref-type="bibr" rid="B7">Li et al. 2024</xref>). Most often, patients choose to be emmetropic at a far distance and use spectacles for near vision. On many occasions, patients need a second pair of glasses for intermediate distances, ranging from 60 cm to 1 m, which can be inconvenient for many patients. In contrast, multifocal <abbrev xlink:title="intraocular lenses">IOLs</abbrev> aim to provide high-quality vision at all three distances without relying on corrective eyewear. However, they may compromise optical quality, which patients experience as halos, glare, and reduced contrast sensitivity (<xref ref-type="bibr" rid="B3">de Silva et al. 2016</xref>; <xref ref-type="bibr" rid="B7">Li et al. 2024</xref>). Extended depth of focus (<abbrev xlink:title="extended depth of focus">EDOF</abbrev>) <abbrev xlink:title="intraocular lenses">IOLs</abbrev> achieve a continuous range of clear vision from intermediate to far distances, reducing the need for glasses compared with traditional monofocal <abbrev xlink:title="intraocular lenses">IOLs</abbrev>. Additionally, they are believed to mitigate the drawbacks of multifocal <abbrev xlink:title="intraocular lenses">IOLs</abbrev>, such as halos and glare (<xref ref-type="bibr" rid="B6">Kohnen et al. 2019</xref>; <xref ref-type="bibr" rid="B4">Fernández et al. 2023</xref>; <xref ref-type="bibr" rid="B10">Sihmar et al. 2023</xref>).</p>
      <p>The Evolux™ enhanced monofocal <abbrev xlink:title="intraocular lens">IOL</abbrev> by the manufacturers AffaMed Therapeutics and SIFI S.p.A. is made from a hydrophobic material and has a non-diffractive profile designed to provide better intermediate vision and equivalent distance clarity compared with a traditional monofocal <abbrev xlink:title="intraocular lens">IOL</abbrev>. The optical technology of the lens provides very low dependence on pupil diameter at far and intermediate distances and similar image contrast at different pupil diameters. Moreover, Evolux™ is reported to be safe in terms of tilt, decentration, and photic phenomena, with good tolerance to residual refractive errors. This study aimed to present the clinical outcomes and patient-reported experiences after implantation of the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> compared with standard monofocal and <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lenses">IOLs</abbrev> (<xref ref-type="bibr" rid="B1">Cano-Ortiz et al. 2024</xref>).</p>
      <p>The novelty of the present study lies in its three-group comparative approach, directly evaluating an enhanced monofocal <abbrev xlink:title="intraocular lens">IOL</abbrev> against both standard monofocal and <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lenses">IOLs</abbrev> in the same clinical setting. In addition to distance and intermediate visual acuity, contrast sensitivity and patient-reported outcomes related to visual satisfaction, spectacle dependence, halos, glare, nighttime visual difficulties, driving difficulties, and diplopia were assessed. This objective and subjective assessment allows the clinical position of the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> to be considered not only in terms of the amount of intermediate vision gained but also in terms of the potential trade-offs associated with achieving that broader range of vision.</p>
    </sec>
    <sec sec-type="Patients and methods" id="sec2">
      <title>Patients and methods</title>
      <p>All patients provided written informed consent for examination and treatment upon hospitalization. The information in this study was obtained through analysis of documentation from the hospitalization and follow-up examinations of the patients. The study protocol was approved by the Ethics Committee of Alexandrovska University Hospital, Sofia (No. 3/23.01.2026). This study adhered to the tenets of the Declaration of Helsinki.</p>
      <p>This retrospective comparative study included 56 patients (67 eyes) who underwent phacoemulsification surgery for cataract with <abbrev xlink:title="intraocular lens">IOL</abbrev> implantation in the capsular bag. The inclusion criteria were uncomplicated cataract surgery, absence of any eye disease other than cataract, absence of postoperative complications, and the ability to read and understand questionnaires. All surgeries were performed by a single experienced surgeon. Patients were excluded if they had corneal astigmatism ≥ −0.50 diopters (<abbrev xlink:title="diopters">D</abbrev>) or had a toric <abbrev xlink:title="intraocular lens">IOL</abbrev> implanted.</p>
      <p>The patients were divided into three groups based on the <abbrev xlink:title="intraocular lens">IOL</abbrev> implanted. The <abbrev xlink:title="intraocular lens">IOL</abbrev> was chosen by the patients after a preoperative discussion with the surgeon. The monofocal <abbrev xlink:title="intraocular lens">IOL</abbrev> group consisted of 25 patients (29 eyes), four of whom underwent bilateral surgery. The monofocal lenses included CT Lucia 621PY by Zeiss, CT Lucia 221P by Zeiss, Alcon Clareon, Synthesis Acrylic monofocal by Cutting Edge, AcrySof IQ SN60WF, and Hoya Vivinex. The Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> group included 17 patients (22 eyes), with five patients undergoing bilateral surgery. The <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lens">IOL</abbrev> group consisted of 14 patients (16 eyes), two of whom underwent bilateral surgery. The <abbrev xlink:title="extended depth of focus">EDOF</abbrev> lenses used were AT LARA 829MP by Zeiss, Synthesis Plus PYPLUS by Cutting Edge, and RayOne EMV by Rayner. The surgical procedures, when both eyes were operated on, were performed 30 days apart.</p>
      <p>All patients underwent preoperative optical biometry using the IOLMaster 700 (Carl Zeiss Meditech, Jena, Germany), and the <abbrev xlink:title="intraocular lens">IOL</abbrev> power was calculated for postsurgical refraction in the range of −0.00 to −0.25 spherical diopters for distance vision. Ophthalmological examination also included best-corrected visual acuity (<abbrev xlink:title="best-corrected visual acuity">BCVA</abbrev>), biomicroscopy, ophthalmoscopy, tonometry, optical coherence tomography (<abbrev xlink:title="optical coherence tomography">OCT</abbrev>), and B-scan ultrasonography.</p>
      <p><abbrev xlink:title="best-corrected visual acuity">BCVA</abbrev> at the far distance was evaluated with the Snellen chart calibrated at 6 m and expressed in decimal notation, where 1.0 (20/20 or 6/6) is normal vision and lower values indicate poorer vision. Visual acuity at the intermediate distance was tested with an intermediate-distance chart calibrated at 66 cm. The chart has a cord attached to it for easier distance measurement.</p>
      <p>Surgery was performed by an experienced surgeon after administering topical anesthesia. Two corneal incisions were made at the 3 and 9 o’clock positions, and one 2.75 mm incision was made at the 12 o’clock position. The incisions were followed by the injection of ophthalmic viscosurgical devices (<abbrev xlink:title="ophthalmic viscosurgical devices">OVDs</abbrev>) using the soft-shell technique, with dispersive and cohesive <abbrev xlink:title="ophthalmic viscosurgical devices">OVDs</abbrev> in the anterior chamber (<abbrev xlink:title="anterior chamber">AC</abbrev>). Curvilinear capsulorrhexis was performed using capsulorrhexis forceps, followed by hydrodissection and phacoemulsification with the horizontal chop technique. The <abbrev xlink:title="intraocular lens">IOL</abbrev> was then implanted into the capsular bag. Patients were excluded from the study if they had intraoperative complications during any stage of the procedure.</p>
      <p>Following cataract surgery, the patients underwent postoperative visits to assess their visual acuity and ophthalmological status on the seventh and 30<sup>th</sup> postoperative days. All patients were administered the same topical therapy, including ofloxacin four times daily during the first week after surgery and a combination of neomycin sulfate/polymyxin B sulfate/dexamethasone 3500 IU/6000 IU/1 mg/mL four times daily for the first week, three times daily for the second week, followed by two times daily for the third week and finally once daily for the fourth week. The examination consisted of postoperative visual acuity (<abbrev xlink:title="visual acuity">VA</abbrev>) without correction at far (6 m) and intermediate (66 cm) distances, biomicroscopy, and ophthalmoscopy. On the 30<sup>th</sup> postoperative day, patients were given a questionnaire regarding their satisfaction with their vision with the operated eye at far and near distances, difficulties in everyday tasks, the appearance of halos, glare, diplopia, and whether they experienced any inconvenience while driving.</p>
      <p>Quantitative variables were analyzed using the Kruskal–Wallis test for comparison of the three groups and post hoc analysis (Dunn–Bonferroni) for pairwise comparisons. The level of statistical significance was set at <italic>p</italic> &lt; 0.05.</p>
    </sec>
    <sec sec-type="Results" id="sec3">
      <title>Results</title>
      <p>In total, 56 patients (67 eyes) were evaluated in this study. The participants had a mean age of 71.80 years (range 54–91 years), of whom 30 were female (53.57%).</p>
      <p>The mean preoperative <abbrev xlink:title="best-corrected visual acuity">BCVA</abbrev> for far vision (6 m) in the monofocal group was 0.18 (ranging from hand movement to 0.5 or 20/40) and 0.25 for intermediate vision at 66 cm (ranging from hand movement to 0.6 or 20/32). For the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> group, the <abbrev xlink:title="best-corrected visual acuity">BCVA</abbrev> for far distance was 0.20 (ranging from PPLC to 0.4 or 20/50) and 0.28 for intermediate distance (ranging from PPLC to 0.6 or 20/32). The <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lens">IOL</abbrev> group showed a mean <abbrev xlink:title="best-corrected visual acuity">BCVA</abbrev> value of 0.21 at 6 m (ranging from PPLC to 0.5 or 20/40) and 0.30 at 66 cm (ranging from PPLC to 0.6 or 20/32). Table <xref ref-type="table" rid="T2">1</xref> presents the results.</p>
      <table-wrap id="T1" position="float" orientation="portrait">
        <label>Table 1.</label>
        <caption>
          <p>Preoperative <abbrev xlink:title="best-corrected visual acuity">BCVA</abbrev> at 6 m and 66 cm.</p>
        </caption>
        <table>
          <tbody>
            <tr>
              <th rowspan="1" colspan="1"/>
              <th rowspan="1" colspan="1">
                <bold>Monofocal <abbrev xlink:title="intraocular lens">IOL</abbrev> group</bold>
              </th>
              <th rowspan="1" colspan="1">
                <bold>Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> group</bold>
              </th>
              <th rowspan="1" colspan="1">
                <bold><abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lens">IOL</abbrev> group</bold>
              </th>
            </tr>
            <tr>
              <td rowspan="1" colspan="1">
                <bold>Mean value of the preoperative <abbrev xlink:title="best-corrected visual acuity">BCVA</abbrev> at 6 m</bold>
              </td>
              <td rowspan="1" colspan="1">0.18</td>
              <td rowspan="1" colspan="1">0.20</td>
              <td rowspan="1" colspan="1">0.21</td>
            </tr>
            <tr>
              <td rowspan="1" colspan="1">
                <bold>Mean value of the preoperative <abbrev xlink:title="best-corrected visual acuity">BCVA</abbrev> at 66 cm</bold>
              </td>
              <td rowspan="1" colspan="1">0.25</td>
              <td rowspan="1" colspan="1">0.28</td>
              <td rowspan="1" colspan="1">0.30</td>
            </tr>
          </tbody>
        </table>
      </table-wrap>
      <p>The recruited patients had no corneal edema, Descemet’s folds, or <abbrev xlink:title="anterior chamber">AC</abbrev> inflammatory reactions on the first postoperative day. <abbrev xlink:title="visual acuity">VA</abbrev> was evaluated again on the seventh and 30<sup>th</sup> days after surgery, and no complications were documented during this period.</p>
      <p>At the first follow-up, the patients in the monofocal <abbrev xlink:title="intraocular lens">IOL</abbrev> group showed a mean visual acuity at 6 m of 0.79, ranging from 0.4 (20/50) to 1.0 (20/20), with 11 of 29 eyes achieving 1.0 (20/20) vision (37.93%). The evaluation of intermediate-distance visual acuity showed a mean value of 0.53, ranging from 0.2 (20/100) to 0.8 (20/25).</p>
      <p>The mean values for the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> group were as follows: 0.80 for far distance, ranging from 0.4 (20/25) to 1.0 (20/20), with eight of 22 eyes reaching 1.0 vision (36.36%), and 0.78 for the intermediate distance of 66 cm, ranging from 0.4 (20/50) to 1.0 (20/20), with six eyes achieving 1.0 vision (27.27%).</p>
      <p>The <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lens">IOL</abbrev> group demonstrated a mean value of 0.81 for uncorrected <abbrev xlink:title="visual acuity">VA</abbrev> at 6 m, ranging from 0.5 (20/40) to 1.0 (20/20), with five of 16 eyes reaching 1.0 (31.25%). The mean value for uncorrected <abbrev xlink:title="visual acuity">VA</abbrev> at 66 cm was 0.83, ranging from 0.4 (20/50) to 1.0 (20/20), with six eyes reaching 1.0 vision (37.50%). Only two eyes had a <abbrev xlink:title="visual acuity">VA</abbrev> of 0.4 (12.50%).</p>
      <p>On the seventh postoperative day, there was a statistically significant difference in distance visual acuity among the three groups (Kruskal–Wallis test, <italic>p</italic> &lt; 0.01). The <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lens">IOL</abbrev> group demonstrated the highest <abbrev xlink:title="visual acuity">VA</abbrev> at both distances. It showed significantly higher <abbrev xlink:title="visual acuity">VA</abbrev> than the monofocal <abbrev xlink:title="intraocular lens">IOL</abbrev> group (<italic>p</italic> &lt; 0.01) and results closer to those of the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> group (<italic>p</italic> &lt; 0.05).</p>
      <p>At the second follow-up, on the 30<sup>th</sup> postoperative day, the patients in all three groups showed improvement in <abbrev xlink:title="visual acuity">VA</abbrev> at both distances, while others showed no improvement. None of the included patients showed worsening on examination.</p>
      <p>The results were as follows:</p>
      <p>The mean <abbrev xlink:title="visual acuity">VA</abbrev> in the monofocal group was 0.80 at 6 m and 0.54 at 66 cm. The Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> group showed a mean <abbrev xlink:title="visual acuity">VA</abbrev> of 0.83 for far distance and 0.79 for intermediate distance. The patients in the <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lens">IOL</abbrev> group demonstrated a mean <abbrev xlink:title="visual acuity">VA</abbrev> of 0.84 at 6 m and 0.93 at 66 cm. The results are presented in Table <xref ref-type="table" rid="T1">2</xref>.</p>
      <table-wrap id="T2" position="float" orientation="portrait">
        <label>Table 2.</label>
        <caption>
          <p>Postoperative visual acuity on the seventh and 30<sup>th</sup> postoperative days.</p>
        </caption>
        <table>
          <tbody>
            <tr>
              <th rowspan="1" colspan="1"/>
              <th rowspan="1" colspan="1">
                <bold>Monofocal <abbrev xlink:title="intraocular lens">IOL</abbrev> group</bold>
              </th>
              <th rowspan="1" colspan="1">
                <bold>Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> group</bold>
              </th>
              <th rowspan="1" colspan="1">
                <bold><abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lens">IOL</abbrev> group</bold>
              </th>
            </tr>
            <tr>
              <td rowspan="1" colspan="1">
                <bold>Mean value of the postoperative <abbrev xlink:title="visual acuity">VA</abbrev> at 6 m on the seventh postoperative day</bold>
              </td>
              <td rowspan="1" colspan="1">0.79</td>
              <td rowspan="1" colspan="1">0.80</td>
              <td rowspan="1" colspan="1">0.81</td>
            </tr>
            <tr>
              <td rowspan="1" colspan="1">
                <bold>Mean value of the postoperative <abbrev xlink:title="visual acuity">VA</abbrev> at 6 m on the 30<sup>th</sup> postoperative day</bold>
              </td>
              <td rowspan="1" colspan="1">0.80</td>
              <td rowspan="1" colspan="1">0.83</td>
              <td rowspan="1" colspan="1">0.84</td>
            </tr>
            <tr>
              <td rowspan="1" colspan="1">
                <bold>Mean value of the postoperative <abbrev xlink:title="visual acuity">VA</abbrev> at 66 cm on the seventh postoperative day</bold>
              </td>
              <td rowspan="1" colspan="1">0.53</td>
              <td rowspan="1" colspan="1">0.78</td>
              <td rowspan="1" colspan="1">0.83</td>
            </tr>
            <tr>
              <td rowspan="1" colspan="1">
                <bold>Mean value of the postoperative <abbrev xlink:title="visual acuity">VA</abbrev> at 66 cm on the 30<sup>th</sup> postoperative day</bold>
              </td>
              <td rowspan="1" colspan="1">0.54</td>
              <td rowspan="1" colspan="1">0.79</td>
              <td rowspan="1" colspan="1">0.93</td>
            </tr>
          </tbody>
        </table>
      </table-wrap>
      <p>On the 30<sup>th</sup> postoperative day, intergroup differences remained statistically significant, although with a smaller magnitude (<italic>p</italic> &lt; 0.05).</p>
      <p><italic>Post hoc</italic> analysis revealed better intermediate <abbrev xlink:title="visual acuity">VA</abbrev> with the <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lens">IOL</abbrev> than with the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> (<italic>p</italic> &lt; 0.05) and the monofocal <abbrev xlink:title="intraocular lenses">IOLs</abbrev> (<italic>p</italic> &lt; 0.001).</p>
      <p>On postoperative day 30, the patients in the monofocal <abbrev xlink:title="intraocular lens">IOL</abbrev> group showed a 5.06% increase in mean <abbrev xlink:title="visual acuity">VA</abbrev> at 6 m and an 8.69% increase in mean <abbrev xlink:title="visual acuity">VA</abbrev> at 66 cm. The percentage increase in the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> group was 3.75% at 6 m and 6.76% at 66 cm. The <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lens">IOL</abbrev> group demonstrated a 3.70% increase in <abbrev xlink:title="visual acuity">VA</abbrev> at 6 m and a 12.05% increase in <abbrev xlink:title="visual acuity">VA</abbrev> at 66 cm, which was the highest increase among all groups. This is likely due to neuroadaptation for intermediate distance occurring over the 30-day period while adjusting to the specific characteristics of the <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lens">IOL</abbrev> optics (<xref ref-type="bibr" rid="B9">Rosa et al. 2017</xref>).</p>
      <p>During the same examination, contrast sensitivity (<abbrev xlink:title="contrast sensitivity">CS</abbrev>) was evaluated using the Pelli-Robson chart. The chart was placed 1 m from the patient. The test was performed using the patient’s best-corrected visual acuity for this distance. The Pelli-Robson score quantifies the subject’s contrast sensitivity on a logarithmic scale ranging from 0.00 to 2.15. Scores of approximately 2.00 log units indicate normal <abbrev xlink:title="contrast sensitivity">CS</abbrev>, while lower scores indicate poorer <abbrev xlink:title="contrast sensitivity">CS</abbrev>.</p>
      <p>Contrast sensitivity in all three groups ranged from 1.55 to 2.00 log units. In the monofocal <abbrev xlink:title="intraocular lens">IOL</abbrev> group, there was only one recording of 1.55 log units of <abbrev xlink:title="contrast sensitivity">CS</abbrev> (3.45%) and one recording of 1.70 log units. Four eyes showed 1.85 log units of <abbrev xlink:title="contrast sensitivity">CS</abbrev> (13.79%), and the remaining 23 eyes demonstrated a <abbrev xlink:title="contrast sensitivity">CS</abbrev> of 2.00 log units (79.31%).</p>
      <p>The patients in the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> group had similar results, with one eye having 1.55 log units (4.55%), two eyes having 1.70 log units (9.09%), five eyes showing 1.85 log units of <abbrev xlink:title="contrast sensitivity">CS</abbrev> (22.73%), and the other 15 eyes recording 2.00 log units (68.18%).</p>
      <p>The results for the <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lens">IOL</abbrev> group were as follows: one eye with 1.55 log units of <abbrev xlink:title="contrast sensitivity">CS</abbrev> (6.25%), two eyes with 1.70 log units (12.50%), nine eyes with 1.85 log units (56.25%), and the remaining four eyes with 2.00 log units of <abbrev xlink:title="contrast sensitivity">CS</abbrev> (25%).</p>
      <p>On the 30<sup>th</sup> postoperative day, patients were asked to complete a questionnaire regarding their satisfaction with their vision after surgery. They were required to evaluate their vision at far and intermediate distances without any optical correction on a scale from 0 to 10, where 0 was “very poor” and 10 was “excellent.” The patients were asked whether they experienced any glare or halos around light sources. In addition, they were required to answer “YES” or “NO” regarding whether they had any difficulties with their vision at night, experienced double vision, or had trouble driving.</p>
      <p>Regarding visual satisfaction among patients in the monofocal <abbrev xlink:title="intraocular lens">IOL</abbrev> group, the values varied from “6” to “10” in the questionnaire. Of the 29 operated eyes, only two patients reported a value of “9” (6.90%), and one patient marked an “8” for distance vision and a “6” for intermediate vision (3.44%). The remaining patients were sufficiently satisfied to report a “10” for both their far and intermediate uncorrected visual acuity. Moreover, one patient reported noticing halos around light sources, and the same patient had difficulties driving at night. One patient stated that they had double vision, and another experienced glare.</p>
      <p>The results in the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> group were similar, with only one patient reporting a “9” for both distances (4.55%), while the others reported a “10.” Regarding halos and glare, only one patient reported experiencing glare, whereas none of the others had any complaints.</p>
      <p>Of the 16 eyes with an <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lens">IOL</abbrev>, one patient marked a value of “9” for <abbrev xlink:title="visual acuity">VA</abbrev> (6.25%), while all others evaluated it as a “10.” For intermediate distance, one patient reported a “6” and another an “8.” Three patients reported photic phenomena, such as glare or halos, when looking at light sources (18.75%).</p>
    </sec>
    <sec sec-type="Discussion" id="sec4">
      <title>Discussion</title>
      <p>The comparative analysis demonstrated excellent post-cataract surgery results in patients who chose the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev>, with excellent <abbrev xlink:title="visual acuity">VA</abbrev> at both evaluated distances. Compared with patients with a monofocal <abbrev xlink:title="intraocular lens">IOL</abbrev>, uncorrected visual acuity after surgery was significantly better with the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> at the intermediate distance and similar at 6 m. Patients with an <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lens">IOL</abbrev> showed comparable results to the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> group, with slightly higher values for intermediate <abbrev xlink:title="visual acuity">VA</abbrev>. Additionally, six patients in the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> group reached a <abbrev xlink:title="visual acuity">VA</abbrev> of 1.0 (20/20) for intermediate vision, while the best result in the monofocal group was 0.8 (20/25). <xref ref-type="bibr" rid="B5">Gabrić et al. (2025)</xref> conducted a study that included 81 patients (116 eyes) and found that the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> demonstrated good distance vision and enhanced intermediate vision without the visual disturbances commonly associated with multifocal or <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lenses">IOLs</abbrev>.</p>
      <p>On postoperative day 30, mean distance <abbrev xlink:title="visual acuity">VA</abbrev> was similar among the three groups, ranging from 0.80 to 0.84. This finding is clinically relevant because improvement in intermediate vision should ideally not occur at the expense of distance visual clarity.</p>
      <p>The difference among the groups was more pronounced at the intermediate distance of 66 cm. On postoperative day 30, the mean intermediate <abbrev xlink:title="visual acuity">VA</abbrev> was 0.54 for the monofocal group, 0.79 for the Evolux™ group, and 0.93 for the <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lens">IOL</abbrev> group. The Evolux™ group demonstrated a substantial improvement over conventional monofocal <abbrev xlink:title="intraocular lenses">IOLs</abbrev>, while the <abbrev xlink:title="extended depth of focus">EDOF</abbrev> group, as expected, achieved the highest <abbrev xlink:title="visual acuity">VA</abbrev> at the intermediate distance. These findings are consistent with the intended optical characteristics of the <abbrev xlink:title="intraocular lens">IOL</abbrev> categories.</p>
      <p>Patients with an <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lens">IOL</abbrev> showed the highest percentage increase in intermediate-distance visual acuity compared with the preoperative values (210%), followed by the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> group (182%). Patients with monofocal <abbrev xlink:title="intraocular lenses">IOLs</abbrev> showed the lowest increase (116%). These results indicate greater spectacle independence for intermediate-distance activities in patients with the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev>, potentially improving their quality of life.</p>
      <p>Visual acuity alone does not fully characterize the quality of vision after implantation of an <abbrev xlink:title="intraocular lens">IOL</abbrev>. Contrast sensitivity is important because optical strategies that extend the depth of focus may alter the distribution of light and consequently affect the ability to distinguish objects of similar luminance, especially under less favorable visual conditions.</p>
      <p>In the present study, <abbrev xlink:title="contrast sensitivity">CS</abbrev> measured using the Pelli-Robson chart was generally preserved in the conventional monofocal and Evolux™ groups. A score of 2.00 log units was recorded in 79.31% of eyes in the monofocal group and 68.18% of eyes in the Evolux™ group. In contrast, only 25% of eyes in the <abbrev xlink:title="extended depth of focus">EDOF</abbrev> group achieved 2.00 log units, while more than half of the <abbrev xlink:title="extended depth of focus">EDOF</abbrev> eyes recorded 1.85 log units. Although the sample size limits definitive conclusions, this distribution suggests that the broader range of vision provided by <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lenses">IOLs</abbrev> may be accompanied by a reduction in contrast sensitivity.</p>
      <p><xref ref-type="bibr" rid="B6">Kohnen et al. (2019)</xref> evaluated the visual performance of an <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lens">IOL</abbrev>. In their analysis, the <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lens">IOL</abbrev> showed lower <abbrev xlink:title="contrast sensitivity">CS</abbrev> than the monofocal lenses, which is consistent with the present data. This could potentially compromise visual clarity in some patients, leading to lower satisfaction.</p>
      <p>An important aspect of the present study was the inclusion of patient-reported outcomes. Despite the small sample, patients in all three groups generally reported high satisfaction with their postoperative vision. In terms of photic phenomena, these were most notable in the <abbrev xlink:title="extended depth of focus">EDOF</abbrev> group according to the questionnaire provided. The lower frequency of photic phenomena in the Evolux™ group is consistent with its non-diffractive enhanced monofocal design. This result further supports the efficacy of the Evolux™ lens, as it offers a lower incidence of halos or glare while providing visual acuity results similar to those of <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lenses">IOLs</abbrev>.</p>
      <p>A similar study on the performance of the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev>, conducted by <xref ref-type="bibr" rid="B1">Cano-Ortiz et al. (2024)</xref>, showed results similar to those of the present study. This observational study included 18 patients and 36 eyes implanted with the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev>. The study showed that the evaluated <abbrev xlink:title="intraocular lens">IOL</abbrev> is a promising solution for cataract surgery, offering remarkable visual acuity at far and intermediate distances with a low incidence of photic phenomena.</p>
      <p>The clinical relevance of the findings is not simply that Evolux™ improves intermediate visual acuity. Rather, the findings suggest a trade-off between the range of vision and visual quality across the three <abbrev xlink:title="intraocular lens">IOL</abbrev> categories. Conventional monofocal <abbrev xlink:title="intraocular lenses">IOLs</abbrev> provided good distance vision but the poorest intermediate performance. <abbrev xlink:title="extended depth of focus">EDOF</abbrev><abbrev xlink:title="intraocular lenses">IOLs</abbrev> provided the highest intermediate acuity but demonstrated lower contrast sensitivity and more frequent photic symptoms. Evolux™ provided intermediate visual clarity between these two strategies, with a substantial gain in intermediate <abbrev xlink:title="visual acuity">VA</abbrev> compared with monofocal lenses while retaining contrast sensitivity and a low frequency of photic complaints.</p>
      <p>This balance may be clinically meaningful when counseling patients about <abbrev xlink:title="intraocular lens">IOL</abbrev> selection. Not every patient requires maximal spectacle independence at all distances. Some patients prioritize excellent distance vision, good intermediate vision, and minimal photic phenomena over achieving the highest <abbrev xlink:title="visual acuity">VA</abbrev> at near distance. For such patients, an enhanced monofocal <abbrev xlink:title="intraocular lens">IOL</abbrev> may represent an attractive alternative to an <abbrev xlink:title="extended depth of focus">EDOF</abbrev> or multifocal <abbrev xlink:title="intraocular lens">IOL</abbrev>.</p>
      <p>The economic aspect may also be relevant to clinical decision-making because enhanced monofocal <abbrev xlink:title="intraocular lenses">IOLs</abbrev> may offer an alternative for patients for whom the additional cost of premium <abbrev xlink:title="intraocular lenses">IOLs</abbrev> is a limiting factor.</p>
      <p>The principal strengths of this study include the direct comparison of three clinically distinct <abbrev xlink:title="intraocular lens">IOL</abbrev> strategies and the assessment of both objective and subjective postoperative outcomes. In particular, combining distance and intermediate visual acuity with contrast sensitivity and patient-reported photic phenomena provides a more comprehensive assessment than <abbrev xlink:title="visual acuity">VA</abbrev> alone. The use of a single experienced surgeon and standardized postoperative treatment also reduced some sources of procedural variability.</p>
    </sec>
    <sec sec-type="Conclusion" id="sec5">
      <title>Conclusion</title>
      <p>The present study offers a comprehensive analysis of objective visual acuity results and patient satisfaction with the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> and provides a broad view of <abbrev xlink:title="intraocular lens">IOL</abbrev> performance. The Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> was shown to be a promising solution for patients undergoing cataract surgery who seek to achieve lower spectacle dependence for daily activities at an intermediate distance. The low incidence of photic phenomena and the results of the contrast sensitivity evaluation indicate that the Evolux™ <abbrev xlink:title="intraocular lens">IOL</abbrev> achieves these results with minimal compromise in the quality of vision.</p>
    </sec>
  </body>
  <back>
    <ref-list>
      <title>References</title>
      <ref id="B1">
        <mixed-citation>Cano-Ortiz A, Sánchez-Ventosa Á, González-Cruces T, Villalba González M, López Pérez MD, Díaz-Ramos JC, Prados Carmona JJ, Tejerina Fernández V, Elies Amat D, Villarrubia A (2024) Evaluation of Evolux™ Intraocular Lenses in Cataract Surgery: Clinical Outcomes and Patient Satisfaction. Journal of Clinical Medicine 13(23): 7404. <ext-link xlink:href="10.3390/jcm13237404" ext-link-type="doi">https://doi.org/10.3390/jcm13237404</ext-link></mixed-citation>
      </ref>
      <ref id="B2">
        <mixed-citation>Dell SJ, Hannan SJ, Venter JA, Teenan D, Hannan NC, Raju D, Berry CW, Kiss HJ, Schallhorn JM (2024) Comparative Analysis of Clinical and Patient-Reported Outcomes of a New Enhanced Monofocal IOL and a Conventional Monofocal IOL. Clinical Ophthalmology 18: 1157–1169. <ext-link xlink:href="10.2147/opth.s456332" ext-link-type="doi">https://doi.org/10.2147/opth.s456332</ext-link></mixed-citation>
      </ref>
      <ref id="B3">
        <mixed-citation>de Silva SR, Evans JR, Kirthi V, Ziaei M, Leyland M (2016) Multifocal versus monofocal intraocular lenses after cataract extraction. The Cochrane Database of Systematic Reviews 12(12): CD003169. <ext-link xlink:href="10.1002/14651858.CD003169.pub4" ext-link-type="doi">https://doi.org/10.1002/14651858.CD003169.pub4</ext-link></mixed-citation>
      </ref>
      <ref id="B4">
        <mixed-citation>Fernández J, Rocha-de-Lossada C, Zamorano-Martín F, Rodríguez-Calvo-de-Mora M, Rodríguez-Vallejo M (2023) Positioning of enhanced monofocal intraocular lenses between conventional monofocal and extended depth of focus lenses: a scoping review. BMC Ophthalmology 23(1): 101. <ext-link xlink:href="10.1186/s12886-023-02844-1" ext-link-type="doi">https://doi.org/10.1186/s12886-023-02844-1</ext-link></mixed-citation>
      </ref>
      <ref id="B5">
        <mixed-citation>Gabrić K, Jagić M, Gabrić I, Bohač M (2025) 12-month clinical outcomes of the Evolux™ extended monofocal intraocular lens following cataract surgery. BMC Ophthalmology 25(1): 524. <ext-link xlink:href="10.1186/s12886-025-04363-7" ext-link-type="doi">https://doi.org/10.1186/s12886-025-04363-7</ext-link></mixed-citation>
      </ref>
      <ref id="B6">
        <mixed-citation>Kohnen T, Böhm M, Hemkeppler E, Schönbrunn S, DeLorenzo N, Petermann K, Herzog M (2019) Visual performance of an extended depth of focus intraocular lens for treatment selection. Eye (Lond) 33(10): 1556–1563. <ext-link xlink:href="10.1038/s41433-019-0443-x" ext-link-type="doi">https://doi.org/10.1038/s41433-019-0443-x</ext-link></mixed-citation>
      </ref>
      <ref id="B7">
        <mixed-citation>Li J, Sun B, Zhang Y, Hao Y, Wang Z, Liu C, Jiang S (2024) Comparative efficacy and safety of all kinds of intraocular lenses in presbyopia-correcting cataract surgery: a systematic review and meta-analysis. BMC Ophthalmology 24(1): 172. <ext-link xlink:href="10.1186/s12886-024-03446-1" ext-link-type="doi">https://doi.org/10.1186/s12886-024-03446-1</ext-link></mixed-citation>
      </ref>
      <ref id="B8">
        <mixed-citation>Moshirfar M, Milner D, Patel BC (2023) Cataract Surgery. [Updated 2023 Jul 18]. In: StatPearls [Internet]. Treasure Island (FL): StatPearls Publishing. <ext-link xlink:href="https://www.ncbi.nlm.nih.gov/books/NBK559253/" ext-link-type="uri">https://www.ncbi.nlm.nih.gov/books/NBK559253/</ext-link></mixed-citation>
      </ref>
      <ref id="B9">
        <mixed-citation>Rosa AM, Miranda ÂC, Patrício MM, McAlinden C, Silva FL, Castelo-Branco M, Murta JN (2017) Functional magnetic resonance imaging to assess neuroadaptation to multifocal intraocular lenses. Journal of Cataract &amp; Refractive Surgery 43(10): 1287–1296. <ext-link xlink:href="10.1016/j.jcrs.2017.07.031" ext-link-type="doi">https://doi.org/10.1016/j.jcrs.2017.07.031</ext-link></mixed-citation>
      </ref>
      <ref id="B10">
        <mixed-citation>Sihmar Y, Singh BK, Singh VK, Kumar S, Chaudhary A, Saini S (2023) Extended depth of focus intraocular lens versus a new monofocal intraocular lens: A prospective comparative and interventional study. Saudi Journal of Ophthalmology 37(4): 331–336. <ext-link xlink:href="10.4103/sjopt.sjopt_68_22" ext-link-type="doi">https://doi.org/10.4103/sjopt.sjopt_68_22</ext-link></mixed-citation>
      </ref>
    </ref-list>
    <sec sec-type="Additional information" id="sec6">
      <title>Additional information</title>
      <sec sec-type="Conflict of interest" id="sec7">
        <title>Conflict of interest</title>
        <p>The authors have declared that no competing interests exist.</p>
      </sec>
      <sec sec-type="Ethical statements" id="sec8">
        <title>Ethical statements</title>
        <p>Clinical trials: The study protocol was approved by the Ethics Committee of Alexandrovska University Hospital, Sofia (No. 3/23.01.2026).</p>
        <p>The authors declared that no experiments on humans or human tissues were performed for the present study.</p>
        <p>Informed consent from the humans, donors or donors’ representatives: Alexandrovska University Hospital</p>
        <p>The authors declared that no experiments on animals were performed for the present study.</p>
        <p>The authors declared that no commercially available immortalised human and animal cell lines were used in the present study.</p>
      </sec>
      <sec sec-type="Artificial Intelligence (AI) use" id="sec9">
        <title>Artificial Intelligence (AI) use</title>
        <p>The authors accept full responsibility for the content of the manuscript, including the disclosure of any use of AI.</p>
        <p>No AI tools were used in the preparation of this manuscript.</p>
      </sec>
      <sec sec-type="Funding" id="sec10">
        <title>Funding</title>
        <p>This study received no external funding.</p>
      </sec>
      <sec sec-type="Author contributions" id="sec11">
        <title>Author contributions</title>
        <p>Yani Zdravkov – methodology, gathered and analyzed the data, and wrote and reviewed the manuscript. Radomir Dzhambazov: Conception and design, gathered research data, analyzed data, and wrote the manuscript. Nevyana Veleva, Vasil Haykin and Alexander Oscar gathered, analyzed and interpreted the data and performed ophthalmological examination, methodology, review, and editing.</p>
      </sec>
      <sec sec-type="Author ORCIDs" id="sec12">
        <title>Author ORCIDs</title>
        <p>N. Veleva <ext-link xlink:href="https://orcid.org/0000-0002-8055-3451" ext-link-type="uri">https://orcid.org/0000-0002-8055-3451</ext-link></p>
        <p>V. Haykin <ext-link xlink:href="https://orcid.org/0000-0001-8539-8860" ext-link-type="uri">https://orcid.org/0000-0001-8539-8860</ext-link></p>
        <p>A. Oscar <ext-link xlink:href="https://orcid.org/0000-0003-3492-9531" ext-link-type="uri">https://orcid.org/0000-0003-3492-9531</ext-link></p>
      </sec>
      <sec sec-type="Data availability" id="sec13">
        <title>Data availability</title>
        <p>All of the data that support the findings of this study are available in the main text.</p>
      </sec>
    </sec>
  </back>
</article>